Probe Station
IV and CV measurements in the Liverpool clean room.
Cold IV:
Setup to measure IV PCB's at temperatures below 0°C. This is done by placing the PCB on a holder with a fan in a commercial freezer. The issue with this system is that the freezer has a cooling cycle and the temperature varies a few degrees when not constantly running. The lowest possible temperature is approximately -23°C. To get to lower temperatures and also have a more stable temperature of the device, a new cold IV setup was designed which will be placed in the freezer. The PCB is cooled with a Peltier element. The hot side is connected with a chuck which allows cooling with a liquid (or gas). The PCB is covered with a Perspex box which will be flushed with Nitrogen to prevent condensation.
TCT PID controller: more information about the PID controller
For IV study: glued sensor on PCB, PT1000 glued on sensor, soldered to PCB
Drawings
Cold IV chuck
Peltier cooled cold IV chuck for PCB; Perspex cover to flush set-up with Nitrogen
Assembled Cold IV setup:
Storage Box
Perspex storage box for PCB's
IV-Box assembled with PCB's:
Additional equipment
nylon screw and screw insulator (also for TCT setup)
Screw types
apt labview vi for IV measurements to controll PID and read temperature
Software
IV/CV standard
V1.3:
standard version
Use LabView 7.1; for newer LabView use new version of programm
Location:
C:\ICV_meas_7.1\V1.3_latest\
V1.3.zip: IV programm V1.3
V1.3.2:
fixed bug in programm flow; 'Abort' button in ICV run window; included switch which allows automatic ramp down at end of measurement; automatic save button opens save window after measurement
Location:
C:\ICV_meas_11
V1.3.2.zip: IV programm V1.3.2
V1.3.3:
ramp down speed available; current limit now user defined number
Location:
C:\ICV_meas_11
V1.3.3.zip: IV programm V1.3.3
Cold IV
Location:
C:\ColdIV
V2.1:
fixed bug in programm flow; 'Abort' button in ICV run window; included switch which allows automatic ramp down at end of measurement; automatic save button opens save window after measurement; ramp down speed available for user; additional read Keithley 2000 multimeter (Ohm) to get resistance of PT1000
V2.1.zip: Cold IV V2.1
V2.2:
current limit now user defined number
V2.2.zip: Cold IV V2.2
--
SvenWonsak - 17 Sep 2014