ATLAS Upgrade: tests of strip modules



Usefull Webpages:

https://twiki.cern.ch/twiki/bin/view/Atlas/SCTUpDAQConfig

https://twiki.cern.ch/twiki/bin/view/Atlas/StripsUpgradeDAQ#Software_installation

Single Hybrid Tests on Panel

Module Tests (with test frame)

In st_system_config need config files for each of the two hybrids like:

  stream ID                           link ID delay hybrid name  
Module 0 1   1 -1 -1 -1 0 0 0 1 1 0 3 0 1 0 50 50 ABC130_Panel_D_Hybrid5_HCC8 ABC130_Test
Module 1 1   1 -1 -1 -1 0 0 0 1 1 0 3 0 3 2 50 50 ABC130_Panel_D_Hybrid2_HCC11 ABC130_Test

the 'stream ID' (first 2 numbers) defines the output stram position in the GUI window

the 'link ID' defines the physical position of the data lines on the connector for the hybrids. Here the left hand hybrid has the ID 3 2 and the right hand hybrid has the ID 1 0

the 'delay' can vary for different systems

for each hybrid a separate config file has to exist with the proper name.

Without Power Board

Hybrid power should be set to ~1.59V with a max current limit of 3 A
  • before configuration the current is approximately:
  • after configuration the current is approximately: 2 A

With Power Board

Hybrid power should be set to ~10.12V with a max current limit of 4 A
  • before configuration the current is approximately: 0.3 A
  • after configuration the current is approximately: 0.41 A