TCT Data Aquisition 
 
 
-   For the DESY analysis program always start voltage scans at 0V and the go to higher voltages (or lower when applying negative bias). 
-   Only use every voltage once when using DESY analysis program  
-   always press rear at power supply (not done automatically) 
  General Settings 
 
-  Byte order: default setting 2 should be used
-  Time between moves: 0.1s recommended
-  DRS capable of taking around 300 wavefroms per second
-  dont highlight whole cell in position moving window when reset position, it will move to the old value, but sets it as 0!
  Laser: 
 
-  For 'Particulars LASER control' only switch tab when laser is OFF, otherwise programm crash -> switch power off and on to controll it again.
-  signal from amplifier very high: reach limit of DRS input 
-  close iris to reduce signal
-  Gregor recommends to have the iris at ~0.8cm (fixed)
-  there is a 'neutral density filter' NDF which attenuates the laser by factor 10
-   when setting the pulse width, don't press 'Disable DAC' after changing to the new value 
 
  HV-CMOS: 
with "Ettore PCB"
position H35DEMO, 2E13 (09/10/2017)
	
		
			|  | PSTCT position | comment | 
		
			| X | 30210 |  | 
		
			| Y | 38000 | not focused | 
		
			| Z | 19590 |  | 
	
positions 24/08/2017
	
		
			|  | PSTCT position | comment | 
		
			| X | 30303 |  | 
		
			| Y | 35995 |  | 
		
			| Z | 19600 |  | 
	
  LFoundry chip (2017) 
positions 22/03/2017
	
		
			|  | PSTCT position | comment | 
		
			| X | 30555 |  | 
		
			| Y | 35800 | after focus scan | 
		
			| Z | 19900 - 20050 | when focussed | 
	
 
-  useing eTCT configuration with HV-CMOS PCB (90 degree angle mount, laser height block)
positions after reset 0, as recommended in Particulars manual
	
		
			|  | PSTCT position | comment | 
		
			| X | 34117.7 |  | 
		
			| Y | 34800 -  37600 | focus position | 
		
			| Z | 19700 - 20100 |  | 
	
positions 21/02/2017
	
		
			|  | PSTCT position | comment | 
		
			| X | 33830 |  | 
		
			| Y | 36000 | after focus scan | 
		
			| Z | 19800 - 20000 | when focussed | 
	
Comment: waveform not good (too long) which affects data quality and focus scan precision; also in the most recent scans the double junction was not fully visible
-- 
SvenWonsak - 26 Nov 2014